• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED MAGEC GENERATION 1.5 SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED MAGEC GENERATION 1.5 SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MC2-4570S
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
An investigation report from an implant retrieval center was received and reported that during examination, the rod failed force testing.The rod was disassembled and substantial debris was seen around the leadscrew, the radial bearing, and the extending bar.Additionally the leadscrew was found to be seized within the extending bar.It is unknown why the rod was removed.No additional information is available.
 
Additional Manufacturer Narrative
The device was returned to newcastle university; no product was returned to the manufacturer for investigation.A root cause was unable to be determined with the information provided.
 
Event or Problem Description
No additional information has been provided.
 
Additional Manufacturer Narrative
A review of the dhr documents indicates the device was manufactured by the specified requirement at the time and met all the required inspections upon release to finished goods.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC GENERATION 1.5 SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED
101 enterprise suite 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED
101 enterprise suite 100
aliso viejo CA 92656
Manufacturer Contact
patrick yrigoyen
101 enterprise suite 100
aliso viejo, CA 92656
8583442667
MDR Report Key19983552
Report Number3006179046-2024-00172
Device Sequence Number2846657
Product Code PGN
UDI-Device Identifier00812258022600
UDI-Public812258022600
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K233593
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 12/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2-4570S
Device Lot Number8092738
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2024
Supplement Date Received by Manufacturer11/19/2024
Initial Report FDA Received Date08/14/2024
Supplement Report FDA Received Date12/16/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/27/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-