MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
SMHX775 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
05/22/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: rupture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation revision surgery with implantation of a 775cc mentor memorygel xtra breast implant prosthesis.Post-operatively, the patient was diagnosed with a ruptured left breast implant using computerized tomography imaging.As a result, the patient underwent unilateral removal and replacement with a left-sided 775cc mentor memorygel xtra breast implant on (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
On september 04, 2024, the mentor analysis lab received the suspect medical device for evaluation.On september 11, 2024, mentor completed an evaluation on the returned device.Mentor conducted a visual inspection and microscopic examination of the device.Visual analysis of the returned sample revealed that the breast implant was found to be ruptured.Additionally, a crease was observed on the edges of the rupture.A microscopic examination was performed, and instrument damage was not found at the edges of the tear.The evaluation determined that the possible cause of the rupture was consistent with normal wear.The instructions for use state that ruptures can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following may cause implants to rupture: stressing the implant during implantation or other procedures and weakening it; folding or wrinkling of the implant shell.Breast implants may also wear over time.Manufacturer¿s reference number: (b)(4).
|
| |
|