|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected fields: b5, d10 and g2.(unmark company representative.) a review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 4 years since the subject device was manufactured.The device was returned to olympus for inspection, and the customer's complaint tip of the channel brush remained in the channel of the scope confirmed.Based on the results of the investigation, it is likely the tip of the channel cleaning brush remained in the channel of the scope that the wire was repeatedly bent for some reason, leading to its breakage.However, a definitive root cause could not be identified.The following is included in the instructions for use (ifu): -after using, carefully check that no parts of the brush have fallen off inside the endoscope's suction channel.If this inspection determines that a part or parts of the brush have come off during cleaning, immediately retrieve them by passing a new instrument or other endotherapy accessory through the channel.Any part of the brush remaining in the channel after cleaning can drop into the patient's body during a subsequent procedure.Depending on the location of a detached part, it may not be recoverable by passing a new instrument or other endotherapy accessory through the endoscope's suction channel.In this case, contact olympus.-clean only one endoscope with this instrument and discard the brush immediately after use.Otherwise, the cleaning effectiveness of the instrument may be impaired, equipment damage and/or infection of the patient and/or operator may occur.-do not withdraw the instrument quickly from the equipment being cleaned.Doing so may cause operator's infection or irritation from the splashes of patient blood/mucus and detergent solution.-exchange this instrument for new one if the brush head is deformed (see figure 4) or the shaft is kinked or deformed (see figure 5).Using a damaged instrument may cause the shaft to be broken, and the broken shaft and/or brush head may remain inside the endoscope¿s channel.-do not withdraw this instrument from the openings of the endoscope at an extreme angle against the insertion direction or with excessive force (see figure 6).Withdraw this instrument slowly with the shaft straight against the insertion direction of the openings of the endoscope (see figure 7,8).Withdrawing the shaft at extreme angel may cause the shaft to be broken and the broken shaft and/or brush head may remain inside the endoscope¿s channel.After withdrawing this instrument from the endoscope, check that there are no abnormities of the brush.Olympus will continue to monitor field performance for this device.
|