The complaint affected unit was not returned for evaluation; therefore, a product non-conformance or device failure could not be confirmed.It is not known if some procedural factors may have contributed to the event.The ifu contains the following warnings: balloon rupture and catheter separation as a result of excessive pull force applied to remove adherent material are the most frequent causes of reported failures.The possibility of balloon rupture must be taken into account when considering the risks involved in any embolectomy procedure; to minimize the risk of vessel damage, balloon rupture, or tip detachment, do not exceed the maximum recommended inflation volume and pull force for each size catheter.Complaints incidence will continue to be monitored, and applicable actions will be taken as required.As part of the manufacturing process 100% of the units go through balloon inspection process.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.Refer to report id 2015691-2024-06140 for the first fogarty catheter that was involved with this event.
|