| Model Number |
G407208 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cardiac Perforation (2513)
|
| Date of Event |
07/23/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product's lot number could not be obtained, therefore the primary di number is provided (d4), but full udi information is not available.
|
| |
|
Event or Problem Description
|
|
During an atrial fibrillation procedure, there was a pericardial effusion.At the beginning of the procedure, the tactisys quartz did not read the contact force, and no reset was possible.When the first tacticath se catheter was connected, prior to insertion into the patient, the contact force information was still unavailable.The tacticath se catheter was exchanged, but the issue was did not resolve.The procedure was then continued with the second tacticath se catheter, despite having no contact force information.Mapping was started with an inquiry afocus mapping catheter but was being finished in the left atrium with the tacticath se catheter when the transseptal access was lost and the tacticath se catheter was back in the right atrium.The patient became hypotensive, and a transesophageal echocardiogram was performed showing a pericardial effusion.The pericardial effusion was noted on the anterior part of the heart between the right and left atrium.No intervention was administered as the pericardial effusion spontaneously resolved on its own.The physician alleges that the perforation occurred during transseptal puncture with the brk transseptal needle or when maneuvering the tacticath se catheter.No ablation was performed.The patient remained stable and was discharged.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d4, g3, h2, h3, h6, h11.The results of the investigation are inconclusive since the device was not returned for analysis.Review of the device history record was not possible as the lot number is unknown.Based on the information received, the cause of the reported pericardial effusion could not be conclusively determined.Per the ifu, cardiac perforation is a known risk during the use of this device.The product's lot number could not be obtained, therefore the primary di number is provided (d4), but full udi information is not available.
|
| |
|
Search Alerts/Recalls
|
|
|