• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ONYX FRONTIER; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ONYX FRONTIER; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number ONYXNG30018UX
Medical Device Problem Codes Deformation Due to Compressive Stress (2889); Device Dislodged or Dislocated (2923); Material Deformation (2976); Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
An attempt was made to use one 2.75 x 26 mm onyx frontier drug eluting stent to treat a moderately calcified lesion in the ostium of the left anterior descending (lad) artery.A 3.0 x 22mm onyx frontier stent and a 3.0 x 18mm onyx frontier stent were also attempted to be used during the procedure to treat a lesion.The devices were inspected before use with no issues noted.The lesion was pre-dilated.The 3.0 x 22mm onyx frontier stent and a 3.0 x 18mm onyx frontier stent were prepped per ifu with no issues noted.The devices did not pass through a previously deployed stent.Resistance was not encountered when advancing the 2.75 x 26 mm onyx frontier stent and excessive force was not used during delivery.It was reported that the 2.75 x 26 mm onyx frontier stent came off the balloon during withdrawal of the device.The stent was being attempted to be deployed but it came off the balloon and had to retrieved out of the vessel by snare.One 3.0 x 22mm onyx frontier stent was also reported to be coming off the balloon but was able to be successfully deployed and implanted in the patient.Intravascular lithotripsy (ivl) was performed first, then the stent came off the non medtronic wire.This caused the plaque to shift and a dissection occurred.The left main was stented back into the aorta to cover a possible spiral dissection.An attempt was made to continue to stent the distal, but five medtronic stents later nothing crossed the lesion.There was no issue removing these five stents from the patient.It was also reported that a stent dislodged of one of these no cross devices, a 3.0 x 18mm onyx frontier stent at an unknown time but believed it may have been after the device was removed from the patient.These five medtronic stents were used after the plaque shift and dissection had occurred.The dissection was not assessed as directly related to the use of the relevant device.The patient is alive with no further injury.The patient is reported to be doing fine.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product analysis summary: the stent was not present on the balloon and did not return for analysis.Kinks were evident to the hypotube.The balloon folds remained intact.Crimp impressions were visible on the exposed balloon surface.Deformation was evident to the distal tip.No other damage evident to the remainder of the device.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONYX FRONTIER
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key19990971
Report Number9612164-2024-03731
Device Sequence Number12503626
Product Code NIQ
UDI-Device Identifier00763000511487
UDI-Public00763000511487
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberONYXNG30018UX
Device Catalogue NumberONYXNG30018UX
Device Lot Number0012217736
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2024
Supplement Date Received by Manufacturer08/28/2024
Initial Report FDA Received Date08/15/2024
Supplement Report FDA Received Date09/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-