| Model Number |
L100 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/16/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable pacemaker exhibited premature battery depletion (pbd).A request was made to have data from this device analyzed.Data analysis confirmed the battery appeared to be depleting more quickly than expected.Also, this device reached elective replacement indicator (eri).Technical services (ts) consultant recommended replacement.This s-icd remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable pacemaker exhibited premature battery depletion (pbd).A request was made to have data from this device analyzed.Data analysis confirmed the battery appeared to be depleting more quickly than expected.Also, this device reached elective replacement indicator (eri).Technical services (ts) consultant recommended replacement.This s-icd remains in service.No adverse patient effects were reported.Additional information indicated that this device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable pacemaker exhibited premature battery depletion (pbd).A request was made to have data from this device analyzed.Data analysis confirmed the battery appeared to be depleting more quickly than expected.Also, this device reached elective replacement indicator (eri).Technical services (ts) consultant recommended replacement.This s-icd remains in service.No adverse patient effects were reported.Additional information indicated that this device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of compromised capacitors.This resulted in the observed premature battery depletion.It was determined that the capacitors were compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of devices in the accolade pacemaker family that have a potential of exhibiting this behavior.This particular device is included in the hydrogen induced premature depletion advisory population.
|
| |
|