MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN MAR LIP LNR 32IDX52OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN MAR LIP LNR 32IDX52OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 121932252
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 08/06/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Patient experiencing pain 4 years post op.No real findings on imagining.Cup removed to review behind the socket.Again, no findings behind the socket of significance.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary patient experiencing pain 4 years post op.No real findings on imagining.Cup removed to review behind the socket.Again, no findings behind the socket of significance.The product and photo of the vent were returned to j&j medtech orthopaedics for evaluation.The medtech orthopaedics team conducted a visual inspection of the returned device and attachment lash revision 6 aug.Jpg.Visual analysis found that the acetabular liner was returned assembled with the acetabular cup, the pinn mar lip lnr 32idx52od presents slightly scratches at the inner surface, most likely caused by the explantation process.However, nothing indicative of a device nonconformance was observed a manufacturing record evaluation was performed for the finished device product description: pinn mar lip lnr 32idx52od product code: 121932252 lot number: j3629d and no non-conformances or manufacturing irregularities were identified.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was not confirmed as the observed condition of the pinn mar lip lnr 32idx52od would have not contributed to the reported adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot a manufacturing record evaluation was performed for the finished device product description: pinn mar lip lnr 32idx52od product code: 121932252 lot number: j3629d and no non-conformances or manufacturing irregularities were identified.Device history review a manufacturing record evaluation was performed for the finished device product description: pinn mar lip lnr 32idx52od product code: 121932252 lot number: j3629d and no non-conformances or manufacturing irregularities were identified.H11 additional narrative: corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR LIP LNR 32IDX52OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19991542
Report Number1818910-2024-17483
Device Sequence Number17904462
Product Code LPH
UDI-Device Identifier10603295014188
UDI-Public(01)10603295014188(17)240430(10)J3629D
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K001534
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Catalogue Number121932252
Device Lot NumberJ3629D
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2024
Supplement Date Received by Manufacturer08/26/2024
10/30/2024
Initial Report FDA Received Date08/15/2024
Supplement Report FDA Received Date08/27/2024
10/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/03/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP; DELTA CER HEAD 12/14 32MM +5; PINN SECTOR W/GRIPTION 52MM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-