| Model Number |
2700TFX23MM |
| Medical Device Problem Code |
Device Stenosis (4066)
|
| Health Effect - Clinical Code |
Heart Failure/Congestive Heart Failure (4446)
|
| Date of Event |
07/22/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
|
| |
|
Event or Problem Description
|
|
It was reported that a 23mm 2700tfx valve in the aortic position underwent a valve-in-valve procedure after an implant duration of 8 years, 4 months due to stenosis.The patient presented with heart failure and dyspnea.The tavr was performed with a 23mm 9755rsl transcatheter valve.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: the device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.A definitive root cause is unable to be determined, however, patient and/or procedural factors likely caused or contributed.
|
| |
|