• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE 132 SIZE 4.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE 132 SIZE 4.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 6020-4535
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Ambulation or Postural Difficulties (2544); Insufficient Information (4580)
Date of Event 07/29/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Patient revised for failure of left total hip arthroplasty.Gray tissue staining.Dissociated head and stem trunnion damage were noted.Stem and head were explanted and replaced with new stem (below billing sheet).Cup and insert were retained.
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation and wear involving a accolade stem was reported.The event was confirmed via review of the provided photographs.Method & results: -product evaluation and results: the device was not returned; however, photographs were provide for review.The photographs show a recently explanted head, some biological material is noted outside and inside the implant.There is also an intraop photo where the stem is shown to be completely disassociated from the stem.Loos of taper lock is confirmed.-clinician review: a review with a clinical consultant indicated " i can confirm that the patient had a primary left total hip arthroplasty with head neck disassociation and trunnion deformity inside was able to review and x-ray showing those findings.I can confirm that the patient had revision surgery by way of the implant usage sheet for the revision procedure although i do not have a copy of the operation report, nor do i have any post revision x-rays.Root cause analysis: the root cause of this event cannot be determined with certainty.Causes of head neck disassociation with trunnionosis are multifactorial including surgical technique especially in the way that the trunnion and femoral head are prepared and implanted, positioning of the implants to avoid impingement, patient factors including lifestyle, activity level, and bmi, as well as implant factors.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to disassociation of the head from the stem and trunnionsosis.The device was not returned; however, photographs were provide for review.The photographs show a recently explanted head, some biological material is noted outside and inside the implant.There is also an intraop photo where the stem is shown to be completely disassociated from the stem.Loos of taper lock is confirmed.A review with a clinical consultant indicated " i can confirm that the patient had a primary left total hip arthroplasty with head neck disassociation and trunnion deformity inside was able to review and x-ray showing those findings.I can confirm that the patient had revision surgery by way of the implant usage sheet for the revision procedure although i do not have a copy of the operation report, nor do i have any post revision x-rays.Root cause analysis: the root cause of this event cannot be determined with certainty.Causes of head neck disassociation with trunnionosis are multifactorial including surgical technique especially in the way that the trunnion and femoral head are prepared and implanted, positioning of the implants to avoid impingement, patient factors including lifestyle, activity level, and bmi, as well as implant factors.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Patient revised for failure of left total hip arthroplasty.Gray tissue staining.Dissociated head and stem trunnion damage were noted.Stem and head were explanted and replaced with new stem (below billing sheet).Cup and insert were retained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE 132 SIZE 4.5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
alessandra chavez
3365 enterprise ave
weston, FL 33331
9546280700
MDR Report Key19994421
Report Number0002249697-2024-01129
Device Sequence Number7057731
Product Code JDI
UDI-Device Identifier04546540510631
UDI-Public04546540510631
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K233498
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2016
Device Catalogue Number6020-4535
Device Lot Number36406801
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/29/2024
Supplement Date Received by Manufacturer07/29/2024
Initial Report FDA Received Date08/15/2024
Supplement Report FDA Received Date09/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/01/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age76 YR
Patient SexMale
-
-