Model Number URF-P7
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The olympus scope was sent to an independent laboratory for culture testing and results are still pending.The investigation is ongoing and follow up with the customer is currently being performed.After culture testing, the device will be evaluated.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event or Problem Description
It was reported that the reno fiberscope tested positive for an unexpected contamination.The issue was found during reprocessing.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the results of third-party testing, the device evaluation and the legal manufacture¿s final investigation.Despite good faith attempts the user culture results and cleaning disinfection and sterilization (cds) processes were not shared.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: (b)(6) 2024.Sampling from: all channels.Cfu: <1cfu.Bacterial identification: n/a.The device was evaluated where no abnormalities were found that could have led to the reported event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the ifu before repair, the results conformed to the regulation's recommendation.The following is included in the device ifu: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was reported that, during reprocessing, the uretero-reno fiberscope tested positive for <1 colony forming units (cfus) of staphylococcus warneri, all channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Common Device Name
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19994454
Report Number9610595-2024-16433
Device Sequence Number7087515
Product Code FGB
UDI-Device Identifier04953170435096
UDI-Public04953170435096
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K181451
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberURF-P7
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2024
Supplement Date Received by Manufacturer10/15/2024
Initial Report FDA Received Date08/15/2024
Supplement Report FDA Received Date10/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/19/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-