• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER INC LATITUDE ELBOW ULNAR CAP MODEL MOYEN; PROSTHESIS, ELBOW, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER INC LATITUDE ELBOW ULNAR CAP MODEL MOYEN; PROSTHESIS, ELBOW, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Catalog Number DKY068
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Osseointegration Problem (3003)
Health Effect - Clinical Codes Unspecified Infection (1930); Inadequate Osseointegration (2646); Implant Pain (4561)
Date of Event 07/22/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
It was reported: the patient underwent right elbow replacement on (b)(6) 2020.The primary implants in the joint became infected, and a revision surgery was performed approximately 3 months later, where several components were exchanged, debridement / specimens collected / lavage.According to the surgeon, the patient has undergone several washouts / debridement procedures of his right elbow joint, since.The patient has significant co-morbidities, which has prevented more substantial surgery.However, the patient has been experiencing significant pain over recent months.Additionally, it was reported there was a loose ulnar component.Decision made to exchange this component, including humeral spool.Primary ulnar component removed (very easily), canal debrided, multiple specimens taken, copious lavage performed, and a longer ulnar stem cemented in place.
 
Additional Manufacturer Narrative
Correction - please refer to h6 device code.The reported event could be confirmed, because even though the implant itself was not returned an x-ray was provided by the reporting facility that confirms the alleged failure mode.The device inspection revealed the following: the provided x-ray was submitted to a stryker physician for review and confirmed ulnar loosening.A functional and dimensional inspection was not possible since the device was not returned.Based on the investigation and medical opinions, the root cause was determined to be patient related issue.The patient has comorbidities that complicate the proper revision of the implant, making it challenging to address any potential infection.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
 
Event or Problem Description
It was reported: the patient underwent right elbow replacement on (b)(6) 2020.The primary implants in the joint became infected, and a revision surgery was performed approximately 3 months later, where several components were exchanged, debridement / specimens collected / lavage.According to the surgeon, the patient has undergone several washouts / debridement procedures of his right elbow joint, since.The patient has significant co-morbidities, which has prevented more substantial surgery.However, the patient has been experiencing significant pain over recent months.Additionally, it was reported there was a loose ulnar component.Decision made to exchange this component, including humeral spool.Primary ulnar component removed (very easily), canal debrided, multiple specimens taken, copious lavage performed, and a longer ulnar stem cemented in place.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATITUDE ELBOW ULNAR CAP MODEL MOYEN
Common Device Name
PROSTHESIS, ELBOW, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
WRIGHT MEDICAL CORK (TORNIER ORTHOPEDICS IRELAND LTD)
harnetts cross
macroom, co. cork NA
EI   NA
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19996791
Report Number0001649390-2024-00372
Device Sequence Number12543476
Product Code JDB
UDI-Device Identifier00846832010912
UDI-Public(01)00846832010912(17)240123(21)2663AU013
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K070787
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date01/23/2024
Device Catalogue NumberDKY068
Device Lot Number2663AU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/22/2024
Supplement Date Received by Manufacturer12/19/2024
Initial Report FDA Received Date08/15/2024
Supplement Report FDA Received Date01/13/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/23/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexMale
-
-