| Catalog Number |
286750100 |
| Medical Device Problem Code |
Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024 during routine incoming inspection of a loaner set at the site it was observed that 80 in.Lbf t-handle was found to be out of drawing specification.There was no known patient or hospital involvement.The product has been quarantined and is available and is being returned to the repair site.No replacement product is required, and no shipping label is required for the product return.All information regarding this event has been reported.This report is for one (1) viper prime torque handle this is report 1 of 1 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d4: udi: as the serial number for the device involved in the event was not provided, the full udi is currently not available.D9, h3, h6: the subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.E3: reporter is a j&j employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that exhibits signs of normal use.No cosmetic defects were observed on the surface.A dimensional inspection was not performed as it is not applicable to the complaint condition.A functional test was performed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the viper prime torque handle would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): a review of the receiving inspection (ri) for viper prime torque handle was conducted identifying that lot number gb150365 was released in one batch.Batch 1: lot qty of (b)(4) units were released on 04 may 2023 with no discrepancies.Supplier: (b)(4).As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|