MAUDE Adverse Event Report: MAQUET SAS MAQUET POWERLED II; LAMP, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS MAQUET POWERLED II; LAMP, SURGICAL Back to Search Results
Model Number ARDPWT209099A
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Event site name: (b)(6).Additional information will be provided following the conclusion of the investigation.
 
Event or Problem Description
On 13th august, 2024 getinge became aware of an issue with one of surgical lights - maquet powerled ii.It was stated the paint was chipping from fork.We decided to report the issue in abundance of caution as any paint particles falling off into sterile field or during procedure may cause contamination.
 
Additional Manufacturer Narrative
Getinge became aware of an issue with one of surgical lights - maquet powerled ii.It was stated the paint was chipping from fork.We decided to report the issue in abundance of caution as any paint particles falling off into sterile field or during procedure may cause contamination.Based on the information collected, it was established that when the event occurred, the surgical light did not meet its specification and in this way the device contributed to event.It is unknown if claimed device was or was not being used for patient treatment or diagnosis when the event took place.A root cause analysis for paint peeling problem was performed by subject matter experts and concluded that all maquet sas products comply with: iec 60601-1 ed.2.0 & ed.3.0 general requirements for basic safety and essential performance.Iec 60601-2-41 particular requirements for the safety of surgical luminaires and luminaire for diagnosis.Paint definition pfc066.This procedure defines maquet sas¿s requirements for all painted parts.Disinfection products test: the aim of these tests is to detect any incompatibility with disinfectant.The paint chip or paint damages are due to: impacts, collisions (abnormal use).The operating manual includes the instructions to pre-position the arms prior to use, in order to prevent damages.To prevent any similar incident, it is recommended to avoid the collisions between devices.Minor paint chip can be repaired with touch up paint, nevertheless, the parts impacted by serious damage must be replaced.It is also recommended to avoid excessive friction during cleaning or inappropriate cleaning products.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
 
Event or Problem Description
Manufacturer's reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET POWERLED II
Common Device Name
LAMP, SURGICAL
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key19998169
Report Number9710055-2024-00471
Device Sequence Number11869949
Product Code FTD
UDI-Device Identifier03700712406189
UDI-Public(01)03700712406189
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARDPWT209099A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2024
Supplement Date Received by Manufacturer10/24/2024
Initial Report FDA Received Date08/16/2024
Supplement Report FDA Received Date10/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/04/2020
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-