• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON INSERT PS NUM4 12MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON INSERT PS NUM4 12MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5532-G-412-E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/25/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reported event: an event regarding infection involving a triathlon insert was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot or sterile lot referenced.Conclusions: it was reported that the patient was revised due to infection.All stryker products sold as sterile are validated to a minimum sterility assurance level sal of 10^-6 in accordance with applicable iso standards.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
I & d and insert-swap.Pt.Presented with a draining surgical wound following mako-tka on (b)(6) 2024.No suspicion of infection as of yet, but dr.Was concerned about the wound draining "this long after surgery".He opted to perform an extensive i&d and swap out the insert.No complaints filed against the components themselves, no further info available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON INSERT PS NUM4 12MM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key20001133
Report Number0002249697-2024-01146
Device Sequence Number11981317
Product Code MBH
UDI-Device Identifier07613327390469
UDI-Public07613327390469
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K173849
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number5532-G-412-E
Device Lot NumberWL5ENY
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/25/2024
Initial Report FDA Received Date08/20/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age62 YR
Patient SexFemale
-
-