| Model Number |
N/A |
| Medical Device Problem Codes |
Fracture (1260); Adverse Event Without Identified Device or Use Problem (2993); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Osteolysis (2377); Insufficient Information (4580); Aseptic loosening (4883)
|
| Date of Event |
06/20/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent a left knee arthroplasty revision of the tibial components due to unknown reasons approximately seventeen (17) months post-operatively.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10 - concomitant devices - persona medial congruent articular surface left 10mm catalog #: 42512100810 lot #: 65548729, persona cruciate retaining standard porous femoral component catalog #: 42502806601 lot #: 65569607, nexgen trabecular metal standard primary patella catalog #: 00587806535 lot #: 65351278.The complainant has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b4, g3, g6, h2, h3, h6, h11.The following sections were corrected: h4.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: a4, b1, b4, b5, b6, b7, g3, g6, h2, h6, h11.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a left knee arthroplasty revision due to tibial aseptic loosening and fracture as well as osteolysis approximately seventeen (17) months post-operatively.Initial operative notes noted no intraoperative complications.Revision operative notes noted the patient was revised to address tibial loosening.During the procedure, the lateral peg of the tibial tray was noted to have broken off and the tibial component could be moved easily.No intraoperative complications were noted.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, g3, g6, h2, h3, h6, h11.Additional medical records were provided and confirmed the reported event.The root cause remains underdetermined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|