| Model Number |
VTICM5_12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/11/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - work order search: no additional similar complaint type events within associated lots were found.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim #(b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation the lens remains implanted.There are multiple medical device reports associated with this patient.This report is 1 of 3 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.Cause of the event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5 - the lens was repositioned.Lens repositioning did not resolved the issue.Lens was exchanged with a longer legth lens.Cause of the event was reported as unknown.Additional information was requested.No further information is available at this time.This file will be re-opened if additional information is provided.Corrected information: d4 - primary unique device identifier (udi) # (b)(4).Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
D6a - implant date: corrected to (b)(6) 2023.H4 - device manufacture date: 22-jun-2023 corrected to 19-jun-2023.Claim #(b)(4).
|
| |
|