H3 product analysis: evaluation determined that the distal bearing is detached.The likely cause of failure was identified as inadequate preventive maintenance.It was also noted that the burr cannot be inserted, laser markings are faded, color ring is faded and scratched and the tube is scratched and worn.Date of event notification: 26-jun-2024.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|