• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABIOMED, INC. IMPELLA CP; TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABIOMED, INC. IMPELLA CP; TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP Back to Search Results
Model Number IMPELLA CP
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Ischemia (1942)
Date of Event 06/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D.6b the exact date of explant is unknown.The impella device was not received from the customer, therefore, investigation of the device was not possible.Should the device or any new information be received, a supplemental manufacturer device report will be filed.As noted in the impella instructions for use: impella cp with smart assist system section: general patient care considerations ¿assess access site for bleeding and hematoma.¿ section: entering cardiac output use of the repositioning sheath and peel-away introducer ¿remove the peel-away introducer completely from the artery over the catheter shaft to prevent trauma and significant bleeding and apply manual pressure above the puncture site.¿ section: potential adverse events (united states) ¿acute renal dysfunction, aortic valve injury, bleeding, cardiogenic shock, cerebral vascular accident/stroke, death, hemolysis, limb ischemia, myocardial infarction, renal failure, thrombocytopenia and cardiac or vascular injury (including ventricular perforation).¿ this report is being filed as part of a retrospective review of historical records.
 
Event or Problem Description
The complainant reported a patient in acute myocardial infarction/cardiogenic shock was implanted with an impella cp device for mechanical circulatory support.While on support, the right foot well perfused despite no pulses.Two days later, the patient had significant bleeding overnight at the impella/groin site.The patient is getting transfused.The patient was found to have the venous sheath almost out and kinked clearly causing the bleed.The physician refloated the sheath which sealed the groin and the impella was explanted.The patient survived the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPELLA CP
Common Device Name
TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP
Manufacturer (Section D)
ABIOMED, INC.
22 cherry hill dr.
danvers MA 01923
Manufacturer (Section G)
ABIOMED, INC.
22 cherry hill dr.
danvers MA 01923
Manufacturer Contact
derek sammarco
MDR Report Key20005632
Report Number1220648-2024-15480
Device Sequence Number7123417
Product Code OZD
UDI-Device Identifier00813502011388
UDI-Public(01)00813502011388(10)1469291(17)220331
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P140003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model NumberIMPELLA CP
Device Catalogue Number0048-0032
Device Lot Number1469291
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2021
Initial Report FDA Received Date08/20/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/24/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-