D.6b the exact date of explant is unknown.The impella device was not received from the customer, therefore, investigation of the device was not possible.Should the device or any new information be received, a supplemental manufacturer device report will be filed.As noted in the impella instructions for use: impella cp with smart assist system section: general patient care considerations ¿assess access site for bleeding and hematoma.¿ section: entering cardiac output use of the repositioning sheath and peel-away introducer ¿remove the peel-away introducer completely from the artery over the catheter shaft to prevent trauma and significant bleeding and apply manual pressure above the puncture site.¿ section: potential adverse events (united states) ¿acute renal dysfunction, aortic valve injury, bleeding, cardiogenic shock, cerebral vascular accident/stroke, death, hemolysis, limb ischemia, myocardial infarction, renal failure, thrombocytopenia and cardiac or vascular injury (including ventricular perforation).¿ this report is being filed as part of a retrospective review of historical records.
|
|
The complainant reported a patient in acute myocardial infarction/cardiogenic shock was implanted with an impella cp device for mechanical circulatory support.While on support, the right foot well perfused despite no pulses.Two days later, the patient had significant bleeding overnight at the impella/groin site.The patient is getting transfused.The patient was found to have the venous sheath almost out and kinked clearly causing the bleed.The physician refloated the sheath which sealed the groin and the impella was explanted.The patient survived the procedure.
|