Model Number TFNT40
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Date of Event 07/18/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lens was returned for evaluation.Solution is dried on the lens.One haptic is broken in the distal tip area.One haptic is split near the gusset area.The lens is cut in half, typical of insertion and removal from the eye.Both halves are adhered to each other by solution.The lens was cleaned for further evaluation.Scratches and scrape marks are observed on the optic.Marks are observed on both sides of the optic that are similar to those left by a surgical instrument used to grasp the lens.Product history records were reviewed, and the documentation indicated the product met release criteria.The file indicates the use of a qualified cartridge and handpiece with a non-qualified viscoelastic.The reported lens damage was observed.The root cause for the reported complaint may be related to a failure to follow the instruction for use (ifu).The viscoelastic indicated is not qualified for the lens/monarch combination used.Company foldable iols are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported that during cataract surgery with intraocular lens (iol) implant procedure, a scratch in the optics was observed.The lens was removed during initial procedure and completed with another lens.Additional information received stating that patient had posterior capsule rupture and patient outcome was recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TORIC TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20005752
Report Number1119421-2024-01602
Device Sequence Number4376191
Product Code MFK
UDI-Device Identifier00380652390583
UDI-Public(01)00380652390583(17)261207(11)211209(10)15316506
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberTFNT40
Device Catalogue NumberTFNT40.230
Device Lot Number15316506
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/22/2024
Initial Report FDA Received Date08/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BIOVISC; MONARCH III IOL CARTRIDGE D; MONARCH III IOL INJECTOR
Outcome Attributed to Adverse Event Other;
Patient Age75 YR
Patient SexMale
-
-