• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRHYTHM TECHNOLOGIES, INC ZEUS SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRHYTHM TECHNOLOGIES, INC ZEUS SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Inaccurate Information (4051)
Health Effect - Clinical Code Unspecified Heart Problem (4454)
Date of Event 07/13/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The zio at device was returned to irhythm, and the clinical data was downloaded.A review of the clinical data found that the patient wore the zio at device for 3 days of the 7-day prescribed wear period.Irhythm became aware of the misclassified arrhythmia while preparing the final report and notified the hcp on day 13.The investigation revealed that the algorithm misclassified the arrhythmia and the certified cardiographic technician (cct) subsequently agreed with the algorithm interpretation.The zio at clinical reference manual states in the ¿indications for use¿ section that ¿the reports are provided for review by the intended user to render a diagnosis based on clinical judgment and experience.¿ this event is being reported per 21cfr 803 as a product problem /malfunction.This report does not constitute an admission by irhythm that the product described in this report has any defects or has malfunctioned.These terms are included in form fda 3500a and are fixed terms for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.
 
Event or Problem Description
The patient experienced an arrhythmia that met medical doctor notification (mdn) requirements that was not communicated during the wear period.The investigation revealed that a preliminary ecg interpretation provided to the physician was misclassified.Following the wear period and while compiling the final report, the interpretation was amended.The hcp was notified immediately, and irhythm learned that the patient received a pacemaker due to the initial false positive interpretation.It is unknown whether the pacemaker will be removed.
 
Additional Manufacturer Narrative
This supplemental report is being submitted in response to the fda notification received on september 26, 2024, concerning missing udi numbers in mdr submissions from october 2023 through december 2024.In response to this issue, irhythm conducted an investigation and identified a system processing error, which has now been resolved.Please see the update in section d4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEUS SYSTEM
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
IRHYTHM TECHNOLOGIES, INC
699 8th st suite 600
san francisco CA 94103
Manufacturer (Section G)
IRHYTHM TECHNOLOGIES, INC
6550 katella avenue, suite 200
cypress CA 90630
Manufacturer Contact
tara mackinnon
699 8th st suite 600
san francisco, CA 94103
6094600577
MDR Report Key20006481
Report Number3007208829-2024-00422
Device Sequence Number19927545
Product Code DQK
UDI-Device Identifier00850043907046
UDI-Public(01)00850043907046
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K222389
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2024
Supplement Date Received by Manufacturer06/02/2025
Initial Report FDA Received Date08/20/2024
Supplement Report FDA Received Date06/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ZIO AT/SN: (B)(6).
Outcome Attributed to Adverse Event Other;
Patient Age84 YR
Patient SexMale
-
-