• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. CIRCULAR MECH XL SEAL 29MM; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. CIRCULAR MECH XL SEAL 29MM; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ECS29B
Medical Device Problem Code Ejection Problem (4009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent 8/20/2024 d4 batch # unk b3: unknown; captured as awareness date attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.Additional information was requested, and the following was obtained: please explain what ¿the handle would not clamp into the stapler.¿ was it difficult to close the device? would the device not fire? were there malformed staples? answer: the detachable anvil would not reconnect to the instrument when inside the body.Surgeon couldn¿t fire the stapler.Had to open a new instrument new instrument performed appropriately.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unknown procedure, the handle would not clamp into the stapler.There were no patient consequences reported.Unknown if the device is available for return.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CIRCULAR MECH XL SEAL 29MM
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key20007059
Report Number3005075853-2024-06149
Device Sequence Number7257682
Product Code GDW
UDI-Device Identifier10705036022930
UDI-Public(01)10705036022930(17)270930(10)X9656L
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K983536
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberECS29B
Device Lot NumberX9656L
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/13/2024
Initial Report FDA Received Date08/20/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-