• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION HYDRATOME RX 44; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION HYDRATOME RX 44; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00583040
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.
 
Event or Problem Description
It was reported to boston scientific corporation that a hydratome rx 44 was used in the biliary during an endoscopic retrograde cholangiopancreatography (ercp) procedure.The exact procedure date was unknown.During the procedure, the cutting wire broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another hydratome rx 44.There were no patient complications reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that a hydratome rx 44 was used in the biliary during an endoscopic retrograde cholangiopancreatography (ercp) procedure.The exact procedure date was unknown.During the procedure, the cutting wire broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another hydratome rx 44.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h11: investigation results: the returned hydratome rx 44 was analyzed, and a visual and microscopic inspection found that the cutting wire was blackened, kinked and broken from the proximal pierce hole.The working length was kinked and torn from the green band to the end of the rx channel.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of cutting wire break was confirmed.The results of the analysis performed on the returned device found that the cutting wire was kinked and broken from the proximal pierce hole and the working length was kinked and torn from the green band to the end of the rx channel.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.The break in the cutting wire could have been generated if there was contact between the device and the scope during energization or if the device exceeded the maximum of voltage during procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wires integrity, causing premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can bend the cutting wire as observed in the product analysis.Working length kinked could be caused by operational factors, such as manipulating the device through difficult anatomy, the used technique for the device manipulation or by the interaction between the device and the scope (hitting against a hard surface), working length torn is typically caused when the user pulls or removes the guidewire through the c channel or by the technique used during loading or removing the guidewire into the device.The investigation findings and all information available concludes the most probable root cause is an adverse event related to procedure.A labeling review was performed, and from the information available, this device was used per the instructions for use (ifu)/product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDRATOME RX 44
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key20007307
Report Number3005099803-2024-03887
Device Sequence Number12577164
Product Code KNS
UDI-Device Identifier08714729755197
UDI-Public08714729755197
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K013153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00583040
Device Catalogue Number8304
Device Lot Number0034188876
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2024
Supplement Date Received by Manufacturer09/09/2024
Initial Report FDA Received Date08/20/2024
Supplement Report FDA Received Date10/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-