• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH PERCUTANEOUS ARM 255MM; FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH PERCUTANEOUS ARM 255MM; FORCEPS Back to Search Results
Catalog Number 398.754
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j employee.H3, h6: the product was not returned to depuy synthes; however, photos were provided for review.The device associated with this report was not returned to depuy synthes for evaluation.The photographs attached were reviewed, however in the images provided no observations pertaining to the nature of the reported event could be identified.The provided evidence was not sufficient to confirm the reported event of unable to assemble/disassemble.Functionality issues cannot be evaluated through a photo investigation.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the photographs attached don't have any evidence to draw a conclusion about the reported event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, multiple items from the synthes colinear clamp set were sent in for repair.It was noticed that the items needed to be replaced due to the items becoming stuck.There was no patient involvement.This report involves one (1) percutaneous arm 255mm.This is report 2 of 4 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCUTANEOUS ARM 255MM
Common Device Name
FORCEPS
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
BACHLER FEINTECH AG
postfach
hölstein 4434
SZ   4434
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key20007884
Report Number8030965-2024-10055
Device Sequence Number17933058
Product Code HTD
UDI-Device Identifier10886982201584
UDI-Public(01)10886982201584
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number398.754
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/22/2024
Initial Report FDA Received Date08/20/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HOHMANN-STYLE ARM 183MM; PELVIC ARM 225MM; SLIDING MECHANISM
Patient SexUnknown
-
-