| Model Number |
AVA10 |
| Medical Device Problem Codes |
Overheating of Device (1437); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Repair request initiated for device with the report of removal difficulty and overheating.No patient impact reported.Repair is being escalated to product event due to reason for the return.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: evaluation determined that the device was overheating and the maximum temperature was measured to be 120.02f.The likely cause of failure was the telescopic operation and the bur lock operation are hard due to inadequate preventive maintenance.It was also noted that there was stamp deterioration and the bur was not installed/attached.B5:date of notification 2024-july-19.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|