| Catalog Number |
6021-2530 |
| Medical Device Problem Codes |
Material Erosion (1214); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Ambulation or Postural Difficulties (2544)
|
| Date of Event |
07/25/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported: "patient is status post revision ltha due to trunnionosis and dissociation (the stem dislocated from the head.The head stayed in the cup).No additional details were provided." a stem, head, and liner were revised.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding disassociation and wear involving an accolade stem was reported.The event was confirmed via review of the provided photographs.Method & results: product evaluation and results: visual inspection: the device was not returned; however, photographs were provide for review.The photographs show a recently explanted stem covered in biological material; the trunnion is noted to be worn.There is also an intraop photo where the stem is shown to be completely disassociated from the stem.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to disassociation of the head from the stem and trunnionosis.Visual inspection: the device was not returned; however, photographs were provide for review.The photographs show a recently explanted stem covered in biological material; the trunnion is noted to be worn.There is also an intraop photo where the stem is shown to be completely disassociated from the stem.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Search Alerts/Recalls
|