| Catalog Number |
151640405 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the tibial insert would not properly lock into tibial base.Needed to open another implant.Unknown surgical delay.Doe: on (b)(6) 2024.Affected side: right knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: it was reported that the tibial insert would not properly lock into tibial base.Needed to open another implant.Unknown surgical delay.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection revealed the locking tab of the attune ps fb insrt sz 4 5mm deformed.The observed damage can be consistent with a component failure caused by excessive forces applied over material, like assembling the device in an off axis position during surgical process.Properly handling and¿attention to the approved use of the device diminishes the risk of failure.As per attune¿ revision knee system fixed bearing surgical technique rev.D, on page 213, the next steps need to be follow for a proper implantation of the attune ps/cr insert with the revision fixed bearing tibial base: "angle the tibial insert posteriorly and slide the posterior tabs into the posterior undercuts of the tibial base.The fixed bearing tibial insert is impacted into place on the tibial base, using the fixed bearing insert impactor.Position the impactor at approximately 60 degrees on the insert so that the notch rests on the anterior edge of the center of the insert.Use a mallet to strike the fixed bearing insert impactor".A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since mating device was not returned for evaluation.However, the reported condition can be confirmed as a difficulty on the assemblance most likely happened as a consequence of the observed damage.The overall complaint was confirmed as the observed condition of the attune ps fb insrt sz 4 5mm would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Event or Problem Description
|
|
Additional information received states that there was no harm to the patient and only a slight delay in getting another implant.
|
| |
|
Event or Problem Description
|
|
Additional information received states that there was no harm to the patient and only a slight delay in getting another implant.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: it was reported that the tibial insert would not properly lock into tibial base.Needed to open another implant.Unknown surgical delay.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection revealed the locking tab of the attune ps fb insrt sz 4 5mm deformed.The observed damage can be consistent with a component failure caused by excessive forces applied over material, like assembling the device in an off axis position during surgical process.Properly handling and¿attention to the approved use of the device diminishes the risk of failure.As per attune¿ revision knee system fixed bearing surgical technique rev.D, on page 213, the next steps need to be follow for a proper implantation of the attune ps/cr insert with the revision fixed bearing tibial base: "angle the tibial insert posteriorly and slide the posterior tabs into the posterior undercuts of the tibial base.The fixed bearing tibial insert is impacted into place on the tibial base, using the fixed bearing insert impactor.Position the impactor at approximately 60 degrees on the insert so that the notch rests on the anterior edge of the center of the insert.Use a mallet to strike the fixed bearing insert impactor".A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since mating device was not returned for evaluation.However, the reported condition can be confirmed as a difficulty on the assemblance most likely happened as a consequence of the observed damage.The overall complaint was confirmed as the observed condition of the attune ps fb insrt sz 4 5mm would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|