| Model Number |
9733560XOM |
| Medical Device Problem Code |
Application Program Freezes or Fails to Launch (4031)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that while on site and turning on the system, the manufacturer representative (rep) was receiving a "no input detected" on the monitor.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d information references the main component of the system.Other relevant device(s) are: product id: 9735672, #: h3 - a manufacturer representative went to the site to service the system.Replaced the cmos battery.Evaluation of the returned battery found the battery had low voltage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|