Zimvie complaint number (b)(4).A1: patient identifier: unknown / not provided a4: patient weight: unknown / not provided d10: concomitant medical product: int3213, osseotite¿ tapered certain¿ implant 3.25 x 13mm, lot: 2018031646 int3211, osseotite¿ tapered certain¿ implant 3.25 x 11.5mm, lot: 2018062302 int3211, osseotite¿ tapered certain¿ implant 3.25 x 11.5mm, lot: 2018062302 a summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.Summary investigation.
|