| Model Number |
BONE REDUCTION FORCEPS, CURVED, POINTED |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 07/22/2024, it was reported by a sales representative via sems(b)(6) that (2) ar-8943-07 bone reduction forceps were bent.This occurred during a case where another set that had forceps that connected was used to proceed.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.One unpackaged ar-8943-07 bone reduction forceps lot number: 6671827 was received for evaluation.Visual evaluation noted that one of the instrument's curved pointed tip was bent.No functional testing was performed because of the damage to the instrument.The most likely cause for the reported failure can be attributed to wear and tear incurred from repeated use.Manufactured date: 27-nov-2018.
|
| |
|