| Model Number |
N/A |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This incident occurred outside the united states (ous).The ous customer reported an observation of an elevated (above reference interval) high-sensitivity troponin i (tnih), lot 104 initial result that was lower (below reference interval) upon retest of the same sample.The initial result was reported to the physician but not questioned.A corrected report was issued.Another sample drawn from the patient approximately 3 hours later resulted low (below reference interval).The lower results were considered correct.There are no allegations of patient harm, changes in treatment, or delays of diagnosis in association with the discordant elevated tnlh result.The interpretation of results section of the instructions for use (ifu) states, "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." siemens is investigating.
|
| |
|
Event or Problem Description
|
|
The customer obtained an elevated (above reference interval) atellica im high-sensitivity troponin i (tnih), lot 104 initial result that was lower (below reference interval) upon retest of the same sample.The initial result was reported to the physician but not questioned.A corrected report was issued.Another sample drawn from the patient approximately 3 hours later resulted low (below reference interval).The lower results were considered correct.There are no allegations of patient harm, changes in treatment, or delays of diagnosis in association with the discordant elevated tnlh result.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial mdr 1219913-2024-00389 was filed on 19-aug-2024.Correction: sections b5, b6, d1 of the initial report filed on 19-aug-2024 incorrectly referenced the atellica im high-sensitivity troponin i (tnih) assay.These sections have been updated to correctly reference the advia centaur high-sensitivity troponin i (tnih).Additional information: siemens completed investigation for an outside of the united states (ous) customer report of an elevated advia centaur high-sensitivity troponin i (tnih), lot 104 initial result on an advia centaur xpt instrument that was lower upon retest of the same sample.Quality control (qc) was in range at the time of testing and o issues were found with customer sample handling.Return of the patient sample for further investigation is not warranted because discordant result was not reproducible, and no other samples were affected.Based on the available information, the cause of the isolated discordant result cannot be determined.Customer has not had any similar issues.The customer is operational, and the reported issue was resolved by routine troubleshooting.The assay's instructions for use (ifu) states the following, under interpretation of results: "results of this assay should always be interpreted in conjunction with the patient's medical history, clinical presentation, and other findings." in section h6 of this report, the investigation finding and conclusion codes were updated based on the investigation results.
|
| |
|
Event or Problem Description
|
|
The customer obtained an elevated (above reference interval) high-sensitivity troponin i (tnih), lot 104 initial result on an advia centaur xpt instrument that was lower (below reference interval) upon retest of the same sample.The initial result was reported to the physician but not questioned.A corrected report was issued.Another sample drawn from the patient approximately 3 hours later resulted low (below reference interval).The lower results were considered correct.There are no allegations of patient harm, changes in treatment, or delays of diagnosis in association with the discordant elevated tnlh result.
|
| |
|
Search Alerts/Recalls
|