• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. Ø7X 30MM BC IF SCRW, VENTED; BIO SOFT TISSUE FIXATN FASTNR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. Ø7X 30MM BC IF SCRW, VENTED; BIO SOFT TISSUE FIXATN FASTNR Back to Search Results
Model Number Ø7X 30MM BC IF SCRW, VENTED
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
It was reported that during an acl reconstruction and lateral tenodisis surgery the screw did not fit on the screwdriver and when inserting the device, the screw broke inside the patient.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the part number ar-2324pslc.The surgeon switched the surgical technique.He did the fixation with a suture anchor, peek swivelock.It was not necessary to do a second surgery.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 complaint allegation is confirmed.Based on the information provided, which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The method of bone preparation was not specified.The most likely cause for the reported failure can be attributed to user error of the device due to improper bone preparation, misalignment insertion, and/or prying/leveraging of the device during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Ø7X 30MM BC IF SCRW, VENTED
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key20017123
Report Number1220246-2024-07185
Device Sequence Number12909196
Product Code MAI
UDI-Device Identifier00888867277649
UDI-Public00888867277649
Combination Product (Y/N)N
Initial Reporter CountryTC
PMA/510(K) Number
K202535
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 06/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model NumberØ7X 30MM BC IF SCRW, VENTED
Device Catalogue NumberAR-4030C-07
Device Lot Number14948132
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2024
Supplement Date Received by Manufacturer06/05/2025
Initial Report FDA Received Date08/21/2024
Supplement Report FDA Received Date06/26/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-