| Catalog Number |
9-VSD-MUSC-014 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Tricuspid Valve Insufficiency/ Regurgitation (4453)
|
| Date of Event |
07/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, a 27mm navitor valve was implanted.There was no difficulty during or after valve implantation.On (b)(6) 2024, the patient had a membranous septum rupture and returned for treatment.It was believed that the cause of the ventricular membranous septum rupture was due to the annulus stent of the valve.During the ventricular septal defect closure, an arterio-venous loop was performed, and a 14mm amplatzer muscular vsd occluder was placed.The 14mm device was too small, and a large tricuspid leak appeared.The tricuspid leak was believed to be possibly due to a break in the chordae following the realization of the av loop.The 14mm device was removed, and the tricuspid regurgitation resolved.The procedure was completed with the placement of a 18mm amplatzer muscular vsd occluder using a 9f amplatzer trevisio intravascular delivery system (atv) for a satisfactory result.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of tricuspid leak was reported.A returned device assessment could not be performed as the device was not returned for analysis.It was indicated that was believed to be possibly due to a break in the chordae following the realization of the av loop.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Based on the information received, the root cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Search Alerts/Recalls
|