Device investigation details: an analysis of the product could not be performed since a physical sample was not received for evaluation.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Note: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.(b)(4).
|
|
It was reported that a patient underwent an ischemic ventricular tachycardia (isvt) ablation procedure with a webster cs and a crista cath and the patient experienced cardiac arrest and hemorrhage that resulted in death.The patients pressure dropped significantly during transesophageal echocardiogram (tee) prior to transeptal puncture.After this was noticed, cardiopulmonary resuscitation (cpr) was administered.The patients' pressure would recover slightly and then fall again - all while going in and out of ventricular tachycardia (vt).Patient was shocked back into normal rhythm when needed.Interventional radiology (ir) was called to check for aortic and retroperitoneal bleed with no findings.Cardio vascular (cv) surgery was called and placed the patient on ecmo.The patient was stable and on ecmo and then shortly after the patient started to code.Cv surgery, ir and eps were all in the room attempting to figure out the problem.It was noticed the patient¿s abdomen was distended.Physicians surgically opened the belly and found the abdomen was full of blood.After a long-standing effort, anesthesia said the pupils were non-reactive and blown.The cause of the bleed remains unknown to all the physicians and family did not want to do an autopsy.No ablation or transeptal puncture had occurred.The physician's opinion on the cause of the adverse event was patient condition.Significant history includes ventricular aneurysm, severe coronary disease and prior to the case an echo showed lv function of 40%.The echo was taken a few days before the procedure.
|