| Model Number |
AS075 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis: evaluation determined that the device was noisy.The likely cause of the failure is identified as detached distal bearing due to inadequate preventive maintenance.It was also noted that unable to insert tool and unable to run the device.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Repair request initiated for device with the report of noise.It was reported that there was no patient involvement.Repair escalated to product event on evaluation due to detached bearing.
|
| |
|
Search Alerts/Recalls
|