• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC. PK CUTTING FORCEPS, 5MM, 33CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. PK CUTTING FORCEPS, 5MM, 33CM Back to Search Results
Model Number PK-CF0533
Medical Device Problem Code Electrical Shorting (2926)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E2/e3: no information available for the occupation of the initial reporter or whether they are a healthcare professional.The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported the subject device sometimes showed a short circuit.The generator was changed and that didn't work; the procedure was completed once the subject device was changed with a similar device.The issue occurred during a gynecology procedure.There were no reports of patient harm.
 
Event or Problem Description
It was reported the subject device sometimes showed a short circuit.The generator was changed and that didn't work; the procedure was completed once the subject device was changed with a similar device.The issue occurred during a gynecology procedure.There were no reports of patient harm.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation and additional information provided by the customer.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was not returned to olympus for inspection; therefore, the reported failure was not confirmed.A root cause could not be established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PK CUTTING FORCEPS, 5MM, 33CM
Common Device Name
PK CUTTING FORCEPS, 5MM, 33CM
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20019943
Report Number3011050570-2024-10305
Device Sequence Number15287844
Product Code GEI
UDI-Device Identifier00821925035867
UDI-Public00821925035867
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K142759
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPK-CF0533
Device Lot NumberFR388723
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/06/2024
Supplement Date Received by Manufacturer09/09/2024
Initial Report FDA Received Date08/21/2024
Supplement Report FDA Received Date09/19/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-