The device was not returned for analysis.A review of the production records and the corrective and preventive actions (capa) database was performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed no indication of a lot specific issue.The investigation determined the reported difficulties and subsequent treatment appear to be related to the operational context of the procedure.It was reported that after stent implantation, the guide wire was noted to be jailed by the stent.The entrapment of the guide wire was likely due to guide wire placement during the procedure.Additionally, it was reported that when attempting to remove the guide wire with a snare, interaction with the stent resulted in the stent being removed with the wire.There is no indication of a product quality issue with respect to manufacture, design, or labeling.The additional device referenced in b5 is filed under a separate medwatch report number.
|