C.R. BARD, INC. (BASD) -3006260740 BROVIAC CV CATHETER, SINGLE-LUMEN, 2.7F; CHRONIC CATHETERS
|
Back to Search Results |
|
| Catalog Number |
0600040 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562); Expulsion (2933)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/02/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that sometimes post a chronic catheter placement, the cuff was allegedly exposed, sutures not adherent.It was further reported that the catheter slid out without any resistance.Reportedly, the tip of catheter was short but not frayed, not beveled.Additionally, removed line was looking short and suspected possibility of broken piece of broviac line retained at the time of line removal.Broken piece was removed by surgery using interventional radiology.The procedure was completed by removing and replacing the line.Current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiry date: 04/2026) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|