• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE VANTAGE WITH SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE VANTAGE WITH SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED3-027-600-40
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 08/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Continuation of d10: product id ped3-027-550-30 (b696119); product type: ; implant date n/a; explant date n/a medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received a report that two pipelines failed to open distally.The patient was undergoing treatment for an unruptured, saccular aneurysm located in the petros segment.The max diameter was 10+mm, and the neck diameter was 5+mm.The landing zone was 5.27mm distal and 5.70mm proximal.The patient's vessel tortuosity was moderate.It was reported that the both devices they both would not open fully distally and in the middle segments of device around a curve.They tried all clinical techniques in trying to to open with no success.They started with a 6x40mm vantage first with no luck getting to open.The device was removed and they then tried a 5.5x30mm vantage, but still had the same issue.The doctor removed the second device as well and canceled the case.The patient did not experience any injury or complications.The devices were prepared according to the instructions for use (ifu).Ancillary devices include a stryker cat 5 guide catheter, phenom 27 microcatheter, and synchro guidewire.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information received reported that both devices had been placed in a vessel bend/curve and failed to open.There was no damage seen to either device, and no cause for the event was determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE VANTAGE WITH SHIELD TECHNOLOGY
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
7635265677
MDR Report Key20027311
Report Number2029214-2024-01440
Device Sequence Number12911710
Product Code OUT
UDI-Device Identifier00847536037182
UDI-Public00847536037182
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P100018
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPED3-027-600-40
Device Catalogue NumberPED3-027-600-40
Device Lot NumberB716111
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2024
Supplement Date Received by Manufacturer08/16/2024
Initial Report FDA Received Date08/22/2024
Supplement Report FDA Received Date08/22/2024
Date Device Manufactured04/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-