| Model Number |
UP210 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
Air Embolism (1697)
|
| Date of Event |
07/26/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the saline supply ran out during surgery, so the nurse was instructed to replace it.After the replacement, air was mixed in when the stopcock was released, causing the saline to flow after a delay of several seconds.The patient developed air embolism after surgery, but recovered and was discharged.Due to the adverse consequences for the patient and the prolonged hospitalization the case is deemed reportable.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.Internal karl storz reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product was inspected, no defect and no abnormality in the device log was found.The most probable root cause is that the user has injected air into the hose connection when changing the fluid.Device evalaiton was completed on september 17,2024 internal karl storz reference number: (b)(4).
|
| |
|