ZIMMER DENTAL ABUT TAPERED ONE-PIECE SCR-V 3.5MM 2MM; DENTAL ABUTMENT Back to Search Results
Catalog Number TAC2
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: patient weight: unknown / not provided.G4: premarket identification: k013227.
 
Event or Problem Description
The tac2 screw broke inside the implant 14 and it was impossible to remove the piece inside, so the implant had to be removed.Procedure completed.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report, b5: describe event or problem, d6: if implanted date.G3: date received by manufacturer, g6: type of report, h1: type of reportable event, h2: follow up type, h3: device evaluated by manufacturer, h6: adverse event problem, h10: additional narrative.Zimvie received one (1) tac2, (abut tapered one-piece scr-v 3.5mm 2mm) for evaluation.Visual evaluation of the as returned device identified the abutment with signs of use.The abutment was identified fractured at the screw.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.This complaint refers to the specific device being investigated for this complaint record.Device history record (dhr) review was completed for the subject lot number 2021100812.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2021100812 for similar events and one (1) other complaint was identified.Review completed utilizing keywords: fracture screw.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are customer error, or patient factors.Therefore, based on the available information, a device malfunction did occur.The reported event was confirmed with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABUT TAPERED ONE-PIECE SCR-V 3.5MM 2MM
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key20029812
Report Number0002023141-2024-02723
Device Sequence Number8674590
Product Code DZE
UDI-Device Identifier00889024018419
UDI-Public(01)00889024018419(17)261021(10)2021100812
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K011028
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 01/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberTAC2
Device Lot Number2021100812
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/22/2024
Supplement Report FDA Received Date01/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/20/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-