MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INCEPTA CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INCEPTA CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) Back to Search Results
Model Number N160
Medical Device Problem Codes High impedance (1291); Delayed Charge Time (2586); Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited a high out of range shock impedance measurement.Technical services (ts) was consulted and discussed stored device information as well as therapy delivery.The crt-d exhibited the elective replacement indicator (eri) battery status due to a charge time of over 15 seconds during capacitor reform.Previous data upload indicated two years of battery longevity remaining.This crt-d remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that there was no conclusive evidence that the observed out-of-range impedance measurements were due to a product performance issue or an inadequate lead-to-device connection.However, intermittent, out-of-range shock lead impedance measurements with no conclusive evidence of a performance issue or an inadequate lead-to-device connection are likely the result of the low-energy test signal utilized for daily automatic or in-clinic commanded shock lead impedance testing.A software update was released in 2015 to further improve consistency of the impedance test results and provide the clinician with additional diagnostic tools, including programmable shock lead impedance alert limits.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited a high out of range shock impedance measurement.Technical services (ts) was consulted and discussed stored device information as well as therapy delivery.The crt-d exhibited the elective replacement indicator (eri) battery status due to a charge time of over 15 seconds during capacitor reform.Previous data upload indicated two years of battery longevity remaining.This crt-d remains in service.No adverse patient effects were reported.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited a high out of range shock impedance measurement.Technical services (ts) was consulted and discussed stored device information as well as therapy delivery.The crt-d exhibited the elective replacement indicator (eri) battery status due to a charge time of over 15 seconds during capacitor reform.Previous data upload indicated two years of battery longevity remaining.This crt-d remains in service.No adverse patient effects were reported.This crt-d was explanted.A new device was implanted.No additional adverse patient effects were reported.This product has not been returned for analysis.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that there was no conclusive evidence that the observed out-of-range impedance measurements were due to a product performance issue or an inadequate lead-to-device connection.However, intermittent, out-of-range shock lead impedance measurements with no conclusive evidence of a performance issue or an inadequate lead-to-device connection are likely the result of the low-energy test signal utilized for daily automatic or in-clinic commanded shock lead impedance testing.A software update was released in 2015 to further improve consistency of the impedance test results and provide the clinician with additional diagnostic tools, including programmable shock lead impedance alert limits.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited a high out of range shock impedance measurement.Technical services (ts) was consulted and discussed stored device information as well as therapy delivery.The crt-d exhibited the elective replacement indicator (eri) battery status due to a charge time of over 15 seconds during capacitor reform.Previous data upload indicated two years of battery longevity remaining.This crt-d remains in service.No adverse patient effects were reported.This crt-d was explanted.A new device was implanted.No additional adverse patient effects were reported.This product has not been returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA CRT-D
Common Device Name
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key20031129
Report Number2124215-2024-52137
Device Sequence Number11529169
Product Code LWP
UDI-Device Identifier00802526480997
UDI-Public00802526480997
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P010012/S255
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/20/2015
Device Model NumberN160
Device Catalogue NumberN160
Device Lot Number933714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2024
Supplement Date Received by Manufacturer08/28/2024
09/05/2024
09/20/2024
Initial Report FDA Received Date08/22/2024
Supplement Report FDA Received Date08/29/2024
09/09/2024
09/23/2024
Date Device Manufactured11/25/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age79 YR
Patient SexMale
-
-