| Model Number |
N160 |
| Medical Device Problem Codes |
High impedance (1291); Delayed Charge Time (2586); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited a high out of range shock impedance measurement.Technical services (ts) was consulted and discussed stored device information as well as therapy delivery.The crt-d exhibited the elective replacement indicator (eri) battery status due to a charge time of over 15 seconds during capacitor reform.Previous data upload indicated two years of battery longevity remaining.This crt-d remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that there was no conclusive evidence that the observed out-of-range impedance measurements were due to a product performance issue or an inadequate lead-to-device connection.However, intermittent, out-of-range shock lead impedance measurements with no conclusive evidence of a performance issue or an inadequate lead-to-device connection are likely the result of the low-energy test signal utilized for daily automatic or in-clinic commanded shock lead impedance testing.A software update was released in 2015 to further improve consistency of the impedance test results and provide the clinician with additional diagnostic tools, including programmable shock lead impedance alert limits.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited a high out of range shock impedance measurement.Technical services (ts) was consulted and discussed stored device information as well as therapy delivery.The crt-d exhibited the elective replacement indicator (eri) battery status due to a charge time of over 15 seconds during capacitor reform.Previous data upload indicated two years of battery longevity remaining.This crt-d remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited a high out of range shock impedance measurement.Technical services (ts) was consulted and discussed stored device information as well as therapy delivery.The crt-d exhibited the elective replacement indicator (eri) battery status due to a charge time of over 15 seconds during capacitor reform.Previous data upload indicated two years of battery longevity remaining.This crt-d remains in service.No adverse patient effects were reported.This crt-d was explanted.A new device was implanted.No additional adverse patient effects were reported.This product has not been returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that there was no conclusive evidence that the observed out-of-range impedance measurements were due to a product performance issue or an inadequate lead-to-device connection.However, intermittent, out-of-range shock lead impedance measurements with no conclusive evidence of a performance issue or an inadequate lead-to-device connection are likely the result of the low-energy test signal utilized for daily automatic or in-clinic commanded shock lead impedance testing.A software update was released in 2015 to further improve consistency of the impedance test results and provide the clinician with additional diagnostic tools, including programmable shock lead impedance alert limits.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited a high out of range shock impedance measurement.Technical services (ts) was consulted and discussed stored device information as well as therapy delivery.The crt-d exhibited the elective replacement indicator (eri) battery status due to a charge time of over 15 seconds during capacitor reform.Previous data upload indicated two years of battery longevity remaining.This crt-d remains in service.No adverse patient effects were reported.This crt-d was explanted.A new device was implanted.No additional adverse patient effects were reported.This product has not been returned for analysis.
|
| |
|