• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG ILESTO 7 VR-T DF-1; ICD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG ILESTO 7 VR-T DF-1; ICD Back to Search Results
Model Number 383578
Medical Device Problem Code Premature End-of-Life Indicator (1480)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2024
Type of Reportable Event Malfunction
Event or Problem Description
This device was explanted due to eos.No adverse patient events were reported.Should additional information become available, this file will be updated.
 
Additional Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Additional Manufacturer Narrative
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.The icd interrogation revealed the eos battery status, detected on august 13, 2024.The memory content of the icd was inspected, revealing a normal current consumption of this device.However, the amount of charge taken from the battery was verified and the battery condition was not as expected.Therefore, the icd was opened and the inner assembly was inspected.The visual inspection of the inner assembly showed no anomalies.In a next step, the electronic module was attached to an external power supply and the eos status was removed with a technical programmer to check the functionality of the electronic module.The measured current consumption was normal and as expected.Analysis of the electronic module revealed no indication of a malfunction.The battery was sent to the manufacturer for a thorough analysis.The manufacturing records of the battery were inspected, documenting that the battery parameters were within specification during the battery manufacturing.No anomalies were documented during the production process, associated with this battery.The battery was subjected to a visual inspection which did not reveal any signs of damage.In a next step, the battery was opened for destructive analysis.The inspection of the inner assembly identified lithium plating, which led to an elevated internal self-depletion and, as a result, to the clinical observation.Please note, this icd is affected by the field safety corrective action, bio-lqc, initiated in march 2021.
 
Event or Problem Description
This device was explanted due to eos.No adverse patient events were reported.Should additional information become available, this file will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILESTO 7 VR-T DF-1
Common Device Name
ICD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
DE  12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key20031607
Report Number1028232-2024-04455
Device Sequence Number3048142
Product Code LWS
UDI-Device Identifier04035479125349
UDI-Public(01)04035479125349(17)170331
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2017
Device Model Number383578
Device Catalogue Number383578
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2024
Supplement Date Received by Manufacturer08/22/2024
05/07/2026
Initial Report FDA Received Date08/22/2024
Supplement Report FDA Received Date08/23/2024
05/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age63 YR
Patient SexFemale
-
-