Catalog Number 71422671
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Medical device site infection (4845)
Date of Event 08/02/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a tka that took place at an unknown time, a revision surgery took place on (b)(6) 2024 to explant the anthem ps hf insert sz 3-4 15mm due to infection.There has been no report of new pain or infection on the patient at this time.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after a tka that took place at an unknown time, a revision surgery took place on (b)(6) 2024 to explant the anthem ps hf insert sz 3-4 15mm due to infection.There has been no report of new pain or infection on the patient at this time.
 
Additional Manufacturer Narrative
The device was not returned for evaluation; therefore, a device analysis could not be performed.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for anthem¿ total knee system revealed a that acute post-surgical wound infection, late deep wound sepsis and/or low-grade synovitis has been identified as a possible adverse effect.Active, local infection or previous intra-articular infections are identified in contraindications for total knee replacement.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.A review of the sterilization records revealed the batch was sterilized within normal parameters.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include loss of sterility during procedure, injury, patient condition and/or medical history.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHEM PS HF INSERT SZ 3-4 15MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key20032102
Report Number1020279-2024-01633
Device Sequence Number3023868
Product Code JWH
UDI-Device Identifier00885556583913
UDI-Public885556583913
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K032295
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71422671
Device Lot Number24AM06703
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2024
Supplement Date Received by Manufacturer08/22/2024
09/20/2024
Initial Report FDA Received Date08/22/2024
Supplement Report FDA Received Date08/26/2024
09/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/10/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-