|
EXACTECH, INC. TRULIANT TIB IMP CRC INSERT SZ 3.5, 17MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-022-51-3517 |
| Medical Device Problem Codes |
Unstable (1667); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Discomfort (2330); Joint Laxity (4526); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/24/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported via legal documentation that approximately 2 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, pain, stiffness, swelling, discomfort, instability, and limited mobility.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
D10: concomitant devices unknown the device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: corrected health effect - clinical code, medical device problem code, and type of investigation.Manufacturer narrative updated: the reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and instability or due to inclusion of the polyethylene in the packaging recall.However, the reported prosthesis wear and instability could not be confirmed and potential contributions of user, or patient-related considerations to the event could not be assessed as the devices were not available for evaluation and no images, radiographs, or relevant clinical information was provided.
|
| |
|
Search Alerts/Recalls
|
|
|