|
Customer service verified that the customer was using the correct kit components, and a replacement pump was sent to the customer.In follow up with a complaint handler on 8/20/2024, the customer stated that her symptoms first appeared on (b)(6) 2024 and was diagnosed with mastitis.She received an iv antibiotic for 48 ours and 2 weeks of oral antibiotics.She has completed the antibiotics, and the mastitis is resolved.Based on the results of our internal investigation (reference number (b)(4)), it cannot be definitively concluded that the pump caused or contributed to the customer¿s mastitis.The estimated incidence of mastitis in lactating women, whether using a breast pump or not, according to published clinical literature can be as high as 33%.In fact, clinical guidelines suggest the use of a breast pump to facilitate withdrawal of breast milk during bouts of mastitis.The complaint rate of mastitis across all reported failures, across all medela breast pumps, is (b)(4) % for the period of january 2013 to august 2017.Mastitis is usually a benign, self-limiting infection with few consequences for the suckling infant.The risk of mastitis is higher among women who have breastfed previously, especially those with a history of mastitis.Riordan & wambach, 4th ed.P.294: breastfeeding and human lactation.However, in this case, the mother¿s mastitis required prompt medical attention for pain relief and prescription antibiotics to avoid progression to overwhelming sepsis.
|