| Model Number |
TV-AB3480-N |
| Medical Device Problem Code |
Lack of Effect (4065)
|
| Health Effect - Clinical Code |
Clinical Endoleak (4843)
|
| Date of Event |
08/06/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient was implanted with the alto stent graft system to treat an abdominal aortic aneurysm (aaa) on (b)(6) 2023.Approximately one (1) year post index procedure follow-up a computed tomography (ct) revealed a type ia endoleak.The patient is currently being monitored while an intervention is being discussed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.H3 other text: device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the type 1a endoleak is unconfirmed.This is not consistent with the reported adverse event/incident.Procedure related harms, device, user, procedure or anatomy relatedness of this complaint could not be determined with the medical records available for review.No procedure related harms were identified.The final patient status was reported as being monitored.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: g3: awareness date has been updated.H6: investigation finding codes: remove code 3233.H6: investigation conclusion codes: remove code 11.
|
| |
|
Search Alerts/Recalls
|