| Model Number |
VICM5 12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/26/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - investigation type 4110: lens work order search - no similar complaint event(s) within associated lots were found.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.The lens was removed and replaced with a different diopter lens and problem was resolved.Cause of the event was reported as other; "a slight refraction error that cannot be called an error.Patient requested lens replacement".Device did not fail to perform.
|
| |
|
Additional Manufacturer Narrative
|
|
Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation - lens was returned in vial in liquid.Visual inspection found no visible damage to lens.Dimensional and functional inspection found the lens to be within specification.Claim# (b)(4).
|
| |
|