• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE INTREPID COAXIAL POLYMER CONICAL DISPOSABLE I/A HANDPIECE; DEVICE, IRRIGATION, OCULAR SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE INTREPID COAXIAL POLYMER CONICAL DISPOSABLE I/A HANDPIECE; DEVICE, IRRIGATION, OCULAR SURGERY Back to Search Results
Catalog Number 8065752145
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
A physician reported that during surgery, an ophthalmic handpiece was clogged.Surgeon used a syringe with water to unblock clog.Surgery was completed on the same day and no patient impact was reported.This complaint is pertaining the one of two reports received from the initial reporter.
 
Additional Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation for the report of occlusion; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The image was not clear or complete.Therefore, the reported issue could not be confirmed; however, the reported lot number was confirmed.A sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown; therefore, specific action with regards to this complaint cannot be taken.All handpieces are 100% visually inspected prior to being released from the manufacturing facility.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTREPID COAXIAL POLYMER CONICAL DISPOSABLE I/A HANDPIECE
Common Device Name
DEVICE, IRRIGATION, OCULAR SURGERY
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20037776
Report Number2523835-2024-01991
Device Sequence Number12365805
Product Code KYG
UDI-Device Identifier00380657521456
UDI-Public(01)00380657521456(17)251231(11)240125(10)16H175(21)NA
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Catalogue Number8065752145
Device Lot Number16H175
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/25/2024
Supplement Date Received by Manufacturer10/25/2024
Initial Report FDA Received Date08/22/2024
Supplement Report FDA Received Date11/19/2024
Date Device Manufactured01/25/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-