Section a2, a4 and a5: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section d6a - implant date: not applicable, as there is no indication that the lens was implanted.Section d6b - explant date: not applicable, as there is no indication that the lens was implanted.Section e1 - first/given name: unknown, as information was requested but not provided.Section e1 - last name: unknown, as information was requested but not provided.Section e1 - email address: unknown, as information was requested but not provided.Section e1 - telephone number: (b)(6).Device evaluation: product testing could not be performed since the device was not returned for evaluation.Therefore, the reported issue could not be verified, and product quality deficiency could not be determined.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order (po) was performed.The search revealed that no additional complaints were received for this po.No escalation required.Conclusion: a product deficiency has not been identified; therefore, no additional corrective actions have been initiated.Attempts have been made to obtain the missing information.However, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
|
It was reported that when using the preloaded monofocal intraocular lens (iol), the doctor was about to insert the iol into the inserter, he pulled too hard causing one of the haptics to break.As a result, the lens was not inserted into the patient's eye; a different lens was implanted.No additional information has been received.
|
|
Additional information and corrected data: after further follow-up, we received the surgeon's details.Section e1 - first/given name: (b)(6).Section e1 - last name: (b)(6).Section e1 - telephone number: (b)(6).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|